Company Name: SUZUKEN
Origin: japan
Trade Mark: SUZUKEN
Model: KENZ Cardico 306
Specifications:
Technical Specification:
■Input:
• Input Impedance:
10MΩ or more
• Leads :
Standard 12 leads and Cabrera lead
Simultaneous or Real time input
• Patient Input:
OnePatient
• Common Mode Rejection:
100 dB or more
• Frequency response:
0.05 – 150Hz within 3dB
• Time constant :
3.2 seconds or more
• Digital Filter:
AC: -30dB or less, 60 or 50Hz
Drift: -3dB, 0.5Hz Muscle: -3dB, 35Hz or 25Hz
• A/D conversion:
12bit
• Sampling rate:
10000 sampling /sec.
• Defibrillator Protection :
to 5000V, 400J
• Patient leakage current:
10µA or less
• Polarization Voltage:
±500mV or more
Display:
• Display System:
5.7 inch STN liquid crystal monochrome display with backlight, 320×240 dots
• ECG Display:
4/12-Lead
■Print :
• Print Mode:
3chx4-row, 3chx4-row+Rhythm
Arrhythmia Monitoring Summary, ANS Test
• Recording Type:
Thermal Array (8 dots/mm)
• Recording Sensitivity:
2.5, 5, 10 OR 20 mm/mV, Auto selection
• Recording Speed: 5, 10,12.5, 25, OR 50 mm/sec
Paper Size: Roll : 80 mm x 30 M
■ Memory :
• Internal Memory :
200 resting ECG
External Inkjet Printer:
USB Interface1 or 2 pages with Landscape/Portrait in A4/Letter
ECG: 12chx1-row or 6chx2-row with Interpretation Result
■Data Communication :
• ECG Data LAN: IEEE 802.3U 100 BASE,TX x 1
• External Memory, Printer or Keyboard: USB 2.0 x 1
Keyboard
Alphanumeric Touch Key in LCD
Function Keys (Membrane Switch)
Operation mode
Auto, Freeze, Pre-check, Post Stress, Manual, Lead - off Alarm
Arrhythmia monitoring mode: 1-15 min, Filing Mode
■ Interpretation :
• Number of Statement : 147 cases
Neonate : 0 – 11 Years Old
Child : 12 – 18 Years Old
Adult : Over 19 Years Old
■Safety :Class I Type CF
Water Proof Protection: Non water proof device(IPXO)
Flammable anaesthetic: Not applicable
Mode of Operation: Continuous operation
■ Power source :
• Rated Power Supply: 100 to 240 VAC, 50 to 60Hz,90VA
• Rechargeable Battery
• Nikel-Metal Hydride battery
• DC 12 V 1500 mAh FDK model: 10HR-AAU
• Rechrage time: 8 hours
• Continuous operation: 2 hours
■Size :
Dimensions: 266mm(W)×173mm(D)×75mm(H)
Wieght Approx.: 1.5kg (including the rechargeable battery)
■Operating Environment :
• Temperature :-10 – 40°C
Humidity : 25 – 95% RH without dew condensation
Pressure : 70 – 106 kPa
■Storage Environment :
• Temperature :-10 – 60°C
• Humidity :
25 – 95% RH without dew condensation
Pressure:70 – 106 kPa
Standard accessories:
• Power Cable
• AC adaptor
• Limp electrode (containing 4 pieces/box)
• Chest electrode (containing 6 pieces/box)
• Patient Cable
• ECG electrode cream 70g
• Operation manual
• CD-R
• ECG recording paper
• Paper spool
Our professional experts will help you.