Company Name: SUZUKEN
Origin: japan
Trade Mark: SUZUKEN
Model: KENZ CARDICO 1215
Specifications:
Technical Specification:
■ Input:
Input Impedance:
Over 10MΩ
Leads :
Standard 12 leads and Cabrera leads
Simultaneous or Real time input
Patient Input:
One Patient
Common Mode Rejection:
Over 100 dB
Frequency response:
0.05 – 150Hz within -3dB to +0.5 dB
Time constant :
Over 3.2 seconds
Digital Filter:
Hum noise, Drift(weak/normal), Muscle : 25/35 Hz
A/D conversion:
13bit (2.44 µV/digit)
Sampling rate:
Input : 1.0000 sampling/sec
Pacemaker detection auxiliary function 10.000 samples/sec
Interpretation/print : 2 msec
Defibrillator Protection :
Up to 5000V, 400J
Patient leakage current:
Less than 10µA
Display :
Display Screen:
Touch Screen 8.5- inche color LCD touch panel with backlight 800*480 dots
ECG Display:
4/6/12 lead
Print :
Print Mode:
1-Analysis report
2- one sheet report
3- Arrhythmia analysis report
4- 3ch+ Arrhythmia lead ECG
5- 6 ch ECG
6- 12 ch ECG
Printer: Thermal Array 8 dots/mm
Sensitivity:
2.5 , 5 ,10 , 20 mm/mV , Auto selection
Speed: 5, 10,12.5 , 25, 50 mm/sec
Paper Size:
Z-fold paper / Roll paper (210mm width)
■ Safety :
Class I Type CF
Water proof protection : Non water proof device (IPXO)
Mode of operation : Continousoperation
■ Power :
AC 110 ~ 240 V 50-60Hz
Rechargeable battery : 12 VDC, Nickel-Metal Hydride battery FDK model: HHR-38AF25GL
Recharge time : 8 hours
Continous operation : 2.5 hours
Recording ECG for 10 Sec in Auto mode with analysis
■ Size :
• Dimensions: 350mm (W) x 305mm (D) x 89 mm (H)
• Weight: Approx. 4 kg
Operating Environment :
Temperature : 10 to 40°C
Humidity : 25 to 95% RH without dew condensation
Pressure : 700 to 1060 hPa
■ Storage Environment :
Temperature : -20 to 60°C
Humidity : 20 to 90% RH without dew condensation
Pressure : 700 – 1060 hPa
Communication port :
• LAN port
• USB port
Memory :
Internal Memory :
1.000 Resting ECG
External USB :
7.200/8GB Resting ECG (option)
■ Data Communication : USB Port x 1
■ Interpretation :
Number of Statement :
147 cases
Neonate : 0 – 11 Years Old
Child : 12 – 18 Years Old
Adalt : Over 19 Years Old
Operation mode :
AUTo ,Freeze , Pre-check ,Post Stress , Manual , Lead-off Alarm , Arrhythmia monitoring mode : 30 sec , 1-15 min , VPC/SVPC summary , Automatic Nervous system Test
Standard Accessories :
- Lemb electrode
- Chest electrode
- Patient cable
- ECG cream
- AC adapter
- CD-R
- ECG recording paper (Z-fold)
Paper spool
Our professional experts will help you.